Tile:

  • Complete Interrupted Tests.

Description:

  • Sometimes a test is interrupted (software or power error) or a user leaves the Test Status Screen before completing all of the test speeds. 
  • When this happens there is a way to complete the test if the original patient is still on the HUMAC NORM or Biodex machine even if the HUMAC Computer is shut down.

Resolution:

  • Return to the Main Menu. Select Patient. Highlight the Patient's name and select Complete Test (top, center of screen). 
  • The HUMAC will jump to the Test Status screen. The completed tests will show Tested. The uncompleted tests will show Error or Not Tested
  • To complete a test, select the appropriate side and row. Press Single Set. Do this until all the untested tests are complete. 
  • Press OK when done to advance to the Reports Screen.